Output voltage (new sensor) 50 nA ± 10 nA / ppm
Response Time (t95) < 7 sec.
Output Linearity Linear
Repeatability 2% of Signal
Maximum Zero Shift 30 ppm NO Equivalent (0 OC...50OC)
Recommended Load (instrument) 10 Ω
Operating Temperature 0 OC...50OC
Storage Temperature Avoid excess heat & cold
Power Consumption < 1mW
Supply Voltage 12 V (Supply voltage via interface board)
Bias Voltage 300 mV
Weight 35g
Accuracy 0…4000 ppm ± 4% or 25 ppmv
4000…5000 ppm ± 8%
Drift <± 3% Signal / Month
Warranty Period 14 months
Pressure Range 800…1200 mBar NOXO 100 Features
Influence of Pressure Atmospheric < 0.1% / 100 mBar
                                                           
											Measuring Gas  < 0.12% / mBar
											
Influence of 
											Humidity                       Non 
											condensing  None
Max. Recommended Storage Time     6 
											months in Canister ?
Note: The sensor can only be used 
											in conjunction with an interface 
											circuit.
This circuit can be supplied upon 
											request.
Our engineers would be pleased to 
											assist when technical help is
necessary.
? Very low pressure influence
? High reliability
? Low power consumption Maintenance 
											free
? Designed using Envitec’s proven 
											know-how in
automotive emissions analyses.






 
										















 
 